Abstract
A new method for the corona discharge ignition in ion mobility spectrometer has been developed. It substantially improves stability and increases device resolution because sample controlled ionization causes a stable flow of ion-exchange processes. The implemented circuit simplifies the design of the proposed ion source and an electronic control circuit. This controlling circuit allows forming a corona discharge without using any additional ignition electrodes.
Keywords
ion mobility spectrometer, corona discharge, ionization source
Full Text:
PDFReferences
Y. P. Raizer, J. E. Allen, V. I. Kisin, Gas Discharge Physics, Springer, 1991.
Discharge ionization source, Patent No. US 6407382, Date of Patent: 06/18/2002
T. W. Carr, Plasma Chromatography, Plenum Press, New York, 1984.
G. A. Eiceman, Z. Karpas, Ion Mobility Spectrometry, CRC Press, Boca Raton, 1993.
M. Tabrizchi, A. Abedi, “A novel electron source for negative ion mobility spectrometry”, International Journal of Mass Spectrometry, vol. 218, pp. 75-85, 2002.
V. S. Pershenkov, A. D. Tremasov, V. V. Belyakov, A. U. Razvalyaev, V. S. Mochkin, “X-ray ion mobility spectrometer”, Microelectronics Reliability, vol. 46, Issue 2-4, pp. 641-644, February 2006.