ADVANCED SAMPLE IONIZATION METHOD IN ION MOBILITY SPECTROMETER
		DOI Number
		
		
					
							10.2298/FUEE1604613B
					
	
		First page
		
		
					
						613
		
	
		Last page
		
		
			
						620
		
	Abstract
A new method for the corona discharge ignition in ion mobility spectrometer has been developed. It substantially improves stability and increases device resolution because sample controlled ionization causes a stable flow of ion-exchange processes. The implemented circuit simplifies the design of the proposed ion source and an electronic control circuit. This controlling circuit allows forming a corona discharge without using any additional ignition electrodes.  
		Keywords
				ion mobility spectrometer, corona discharge, ionization source
			
		Full Text:
PDFReferences
Y. P. Raizer, J. E. Allen, V. I. Kisin, Gas Discharge Physics, Springer, 1991.
Discharge ionization source, Patent No. US 6407382, Date of Patent: 06/18/2002
T. W. Carr, Plasma Chromatography, Plenum Press, New York, 1984.
G. A. Eiceman, Z. Karpas, Ion Mobility Spectrometry, CRC Press, Boca Raton, 1993.
M. Tabrizchi, A. Abedi, “A novel electron source for negative ion mobility spectrometry”, International Journal of Mass Spectrometry, vol. 218, pp. 75-85, 2002.
V. S. Pershenkov, A. D. Tremasov, V. V. Belyakov, A. U. Razvalyaev, V. S. Mochkin, “X-ray ion mobility spectrometer”, Microelectronics Reliability, vol. 46, Issue 2-4, pp. 641-644, February 2006.
Refbacks
- There are currently no refbacks.
ISSN: 0353-3670 (Print)
ISSN: 2217-5997 (Online)
COBISS.SR-ID 12826626